Standard chest CT using combined automated tube potential selection and iterative reconstruction: image quality and
In Hye Chae1, Yookyung Kim1, So Won Lee1
1Department of Radiology, School of Medicine, Ewha Womans University, Seoul, South Korea.
Objective:
To evaluate the image quality and radiation dose reduction of chest computed tomography (CT) using combined automated tube potential selection (ATPS) and iterative reconstruction (IR).
Materials And Methods:
Image quality and radiation dose were compared between conventional contrast-enhanced chest CT using 120 kVp and filtered back projection (Protocol A) and CT using ATPS and IR (Protocol B) in 43 patients.
Results:
The effective radiation dose was 3.6 ± 0.4 mSv for Protocol A and 2.2 ± 0.6 mSv for Protocol B (mean dose reduction, 39.7%). Protocol B showed diagnostic image quality in all patients.
Conclusion:
ATPS and IR allows for radiation dose reduction while maintaining diagnostic image quality.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography