Objectives:
This study assessed image quality and radiation dose of multidetector computed tomography (CT) examination using a standard protocol and a low-voltage protocol.
Methods:
Patients requiring contrast-enhanced abdominal CT examination were randomly assigned to two groups with different voltage protocols: (i) 120 kV; (ii) an automated attenuation-based tube potential optimization mode (CARE kV). The volume CT dose index (CTDIvol) and dose length product (DLP) were recorded. Image quality was semiquantitatively assessed by two blinded radiologists using a five-point scale.
Results:
There were 39 patients in the 120 kV group and 50 patients in the CARE kV group. There was no obvious difference in image quality score between the groups. CARE kV resulted in a voltage reduction to 100 kV in 45 patients and to 80 kV in five patients. CTDIvol and DLP were significantly lower with CARE kV than with the 120 kV protocol.
Conclusions:
The use of CARE kV reduces radiation dose with no loss of image quality compared with a standard 120 kV protocol.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
X-ray Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...