Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
817
Optimizing precision in toric lens selection by combining keratometry techniques.
Summary
Averaging measurements from manual and automated keratometers significantly reduces outliers in preoperative assessments. This improves the precision of toric intraocular lens selection for cataract surgery.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Surgical Technology
Background:
- Toric intraocular lens (IOL) selection relies on precise preoperative measurements.
- Variability in keratometry technology can lead to outliers and affect IOL accuracy.
- Routine cataract surgery demands reliable methods for accurate toric IOL power calculation.
Purpose of the Study:
- To enhance the precision of toric intraocular lens selection.
- To minimize outliers caused by technology-dependent variability in preoperative assessments.
- To evaluate the impact of averaging measurements from different keratometers on measurement accuracy.
Main Methods:
- Compared preoperative measurements (sphere, astigmatism, axis) from a manual and four automated keratometers for 87 eyes.
- Defined a 'meld' value by averaging measurements across all five technologies for each parameter.
- Utilized Bland-Altman analysis, paired t-tests, and correlation coefficients to assess technology agreement with the meld, quantifying outliers.
Main Results:
- While statistical tests showed insignificant differences between individual technologies and the meld, precision analysis revealed clinically significant outliers.
- Averaging measurements from manual and automated keratometers reduced outliers for sphere (2%-46% to 1%-6%), astigmatism (6%-23% to 0%-2%), and axis (15%-27% to 3%-6%).
- Individual keratometry technologies produced a variable number of outliers, which could be overlooked when using a single device.
Conclusions:
- Multiple keratometry technologies yield similar average measurements but can produce overlooked outliers when used in isolation.
- Averaging measurements from manual keratometry with automated technologies significantly reduces measurement errors.
- This combined approach enhances the precision of toric intraocular lens selection in cataract surgery.