Neonatal resuscitation in resource-limited settings: titrating oxygen delivery without an oxygen blender
Marta Thio1, Liselotte van Kempen2, Anthony R Rafferty3
1Neonatal Services, The Royal Women's Hospital, Melbourne, Australia; Hospital Sant Joan de Deu, Barcelona, Spain; The University of Melbourne, Victoria, Australia.
Objective:
To test whether 4 commonly used self-inflating bags with a reservoir in situ can reliably deliver different oxygen concentrations (21%-100%) using a portable oxygen cylinder with flows of ≤5 L/min.
Study Design:
Four self-inflating bags (from Laerdal, Ambu, Parker Healthcare, and Mayo Healthcare) were tested to provide positive pressure ventilation to a manikin at 60 inflations/min by 4 operators. Oxygen delivery was measured for 2 minutes, combining oxygen flows (0.25, 0.5, 1, 5 L/min) and peak inspiratory pressures (PIPs 20-25, 35-40 cmH2O).
Results:
Combinations (n=128) were performed twice. Oxygen delivery depended upon device, oxygen flow, and PIP. All self-inflating bags delivered mean oxygen concentrations of <40% with 0.25 L/min, regardless of PIP. Three self-inflating bags delivered ≤40% with flow 0.5 L/min at PIP 35-40 cmH2O, whereas all delivered >40% at PIP 20-25 cmH2O. With 1 L/min, 3 self-inflating bags delivered 40%-60% at PIP 35-40 cmH2O and all delivered >60% at PIP 20-25 cmH2O. With 5 L/min, all self-inflating bags delivered close to or 100%, regardless of PIP. Differences in oxygen delivery between self-inflating bags were statistically significant (P<.001) even when differences were not clinically important.
Conclusion:
Self-inflating bags with a reservoir in situ can deliver a variety of oxygen concentrations without a blender, from <40% with 0.25 L/min oxygen flow to 100% with 5 L/min. The adjustment of oxygen flow may be a useful method of titrating oxygen in settings where air-oxygen blenders are unavailable.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...