In vitro evaluation of three intravenous fluid line warmers
Rebecca A Lee1, Heather A Towle Millard, Ann B Weil
1Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Purdue University, West Lafayette, IN 47907.
Objective:
To determine in vitro output temperature differences of 3 IV fluid warmers.
Design:
Prospective, randomized study.
Sample:
3 IV fluid warmers.
Procedures:
Warming capabilities of a distance-dependent blood and fluid warmer marketed for human and veterinary use (product A) and a veterinary-specific distance-dependent fluid warmer (product B) were compared at 0, 4, 8, and 12 cm from the device to the test vein and at flow rates of 20, 60, 100, 140, 180, 220, 260, and 300 mL/h with room temperature (approx 22°C) fluids (phase 1). The superior warming device was compared against a distance-independent IV fluid warmer (product C) with room temperature fluids at the same flow rates (phase 2). The effect of prewarmed fluids (38°C) versus room temperature fluids was evaluated with the superior warming device from phase 2 (phase 3).
Results:
In phase 1, product B produced significantly warmer fluids than product A for all flow rates and distances. Both distance-dependent devices produced warmer fluid at 0 cm, compared with 4, 8, and 12 cm. In phase 2, product B produced warmer fluid than product C at 60, 100, 140, and 180 mL/h. In phase 3, there was no significant benefit to use of prewarmed fluids versus room temperature fluids. Output temperatures ≥ 36.4°C were achieved for all rates ≥ 60 mL/h.
Conclusions And Clinical Relevance:
Product B had superior warming capabilities. Placing the fluid warmer close to the patient is recommended. Use of prewarmed fluids had no benefit. Lower IV fluid flow rates resulted in lower output fluid temperatures.
More Related Videos
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
09:35Modified MicroSecure Vitrification: A Safe, Simple and Highly Effective Cryopreservation Procedure for Human Blastocysts
Published on: March 2, 2017
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...