Motion as perturbation. II. Development of the method for dosimetric analysis of motion effects with fixed-gantry
Benjamin E Nelms1, Daniel Opp2, Geoffrey Zhang2
1Canis Lupus LLC, Merrimac, Wisconsin 53561.
Purpose:
In this work, the feasibility of implementing a motion-perturbation approach to accurately estimate volumetric dose in the presence of organ motion--previously demonstrated for VMAT--is studied for static gantry IMRT. The method's accuracy is improved for the voxels that have very low planned dose but acquire appreciable dose due to motion. The study describes the modified algorithm and its experimental validation and provides an example of a clinical application.
Methods:
A contoured region-of-interest is propagated according to the predefined motion kernel throughout time-resolved 4D phantom dose grids. This timed series of 3D dose grids is produced by the measurement-guided dose reconstruction algorithm, based on an irradiation of a static ARCCHECK (AC) helical dosimeter array (Sun Nuclear Corp., Melbourne, FL). Each moving voxel collects dose over the dynamic simulation. The difference in dose-to-moving voxel vs dose-to-static voxel in-phantom forms the basis of a motion perturbation correction that is applied to the corresponding voxel in the patient dataset. A new method to synchronize the accelerator and dosimeter clocks, applicable to fixed-gantry IMRT, was developed. Refinements to the algorithm account for the excursion of low dose voxels into high dose regions, causing appreciable dose increase due to motion (LDVE correction). For experimental validation, four plans using TG-119 structure sets and objectives were produced using segmented IMRT direct machine parameters optimization in Pinnacle treatment planning system (v. 9.6, Philips Radiation Oncology Systems, Fitchburg, WI). All beams were delivered with the gantry angle of 0°. Each beam was delivered three times: (1) to the static AC centered on the room lasers; (2) to a static phantom containing a MAPCHECK2 (MC2) planar diode array dosimeter (Sun Nuclear); and (3) to the moving MC2 phantom. The motion trajectory was an ellipse in the IEC XY plane, with 3 and 1.5 cm axes. The period was 5 s, with the resulting average motion speed of 1.45 cm/s. The motion-perturbed high resolution (2 mm voxel) volumetric dose grids on the MC2 phantom were generated for each beam. From each grid, a coronal dose plane at the detector level was extracted and compared to the corresponding moving MC2 measurement, using gamma analysis with both global (G) and local (L) dose-error normalization.
Results:
Using the TG-119 criteria of (3%G/3 mm), per beam average gamma analysis passing rates exceeded 95% in all cases. No individual beam had a passing rate below 91%. LDVE correction eliminated systematic disagreement patterns at the beams' aperture edges. In a representative example, application of LDVE correction improved (2%L/2 mm) gamma analysis passing rate for an IMRT beam from 74% to 98%.
Conclusions:
The effect of motion on the moving region-of-interest IMRT dose can be estimated with a standard, static phantom QA measurement, provided the motion characteristics are independently known from 4D CT or otherwise. The motion-perturbed absolute dose estimates were validated by the direct planar diode array measurements, and were found to reliably agree with them in a homogeneous phantom.
More Related Videos
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Absolute Motion Analysis- General Plane Motion
As the drone's propellers rotate, an upward force is generated that counteracts the force of gravity, enabling the drone to lift off from the ground. This initial movement of the drone is along a straight path, representing a form of translational motion. In this phase, every point on the...
Relative Motion Analysis using Rotating Axes
However, to express the relative position of point B relative to point A, an additional frame of reference, denoted as x'y', is necessary. This additional frame not only translates but also rotates relative to the fixed frame, making it...
Equation of Motion for a Rigid Body
The combined moments generated about the center of mass of the object are equal to the rate of change of the angular momentum of the body. An external force, when applied at a different...
Relative Motion Analysis using Rotating Axes-Problem Solving
Here, in order to determine the magnitude of velocity and acceleration for point...
Relative Motion Analysis - Acceleration
Planar Rigid-Body Motion
Planar motion is typically divided into three distinct categories. The first is rectilinear translation, demonstrated by a subway train that moves along...