A method for multichannel dosimetry with EBT3 radiochromic films
Jose Fernando Pérez Azorín1, Luis Isaac Ramos García2, Josep M Martí-Climent3
1Hospital Universitario de Cruces, Department of Medical Physics, Plz. Cruces s/n, Barakaldo, 48903 Bizkaia, Spain.
Purpose:
An improved method for multichannel dosimetry is presented. This method explicitly takes into account the information provided by the unexposed image of the film.
Methods:
The method calculates the dose by applying a couple of perturbations to the scanned dose, one dependent and the other independent on the color channel. The method has been compared with previous multichannel and two single channel methods (red and green) against measurements using two different tests: first, five percentage depth dose profiles covering a wide range of doses; second, the dose map perpendicular to the beam axis for a 15 × 15 cm(2) square field. Finally, the results of 30 IMRT quality assurances tests are presented. All tests have been evaluated using the gamma analysis.
Results:
The coefficient of variation was found to be similar for all methods in a wide range of doses. The results of the proposed method are more in agreement with the experimental measurements and with the treatment planning system. Furthermore, the differences in the mean gamma pass rates are statistically significant.
Conclusions:
The improved multichannel dosimetric method is able to remove many of the common disturbances usually present in radiochromic films and improves the gamma analysis results compared with the other three methods.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Biological Effects of Radiation
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET