Comparison of cyclic fatigue resistance of original and counterfeit rotary instruments
Huseyin Ertas1, Ismail Davut Capar, Hakan Arslan
1Department of Endodontics, Faculty of Dentistry, Izmir Katip Celebi University, Izmir 35620, Turkey. dentisthuseyin@yahoo.com.
Introduction:
In recent years, with the advances in counterfeiting methods, counterfeit products have reached the dental market. The purpose of this study was to compare the cyclic fatigue resistance of original and counterfeit rotary root canal instruments.
Materials And Methods:
The cyclic fatigue of original and counterfeit ProTaper F2 endodontic instruments was tested (n = 20) in 3 mm radius steel canals with a 60° angle of curvature. The number of cycles to fracture (NCF) was calculated, and the data were subjected to the Student's t-test (α = 0.05).
Results:
The original instruments showed better cyclic fatigue resistance than the counterfeit ones (p < .001). The mean NCF was 483 ± 84 for the original files and 186 ± 86 for the counterfeit files.
Conclusions:
The cyclic fatigue resistance of the counterfeit instruments was very low. As a result, clinicians should be careful not to purchase counterfeit products.
More Related Videos
08:47Mimicking and Measuring Occlusal Erosive Tooth Wear with the "Rub&Roll" and Non-contact Profilometry
Published on: February 2, 2018
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Related Concept Videos
Fatigue
Residual Stresses in Circular Shafts
Stress Concentrations in Circular Shafts
Deformation in a Circular Shaft
Rolling Resistance: Problem Solving
Thin-Walled Hollow Shafts