Related Experiment Video
Updated: Apr 28, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Analysis of dose measurement other than the radiation protection during the radiographic examination
Ki-Youl Kim1, Jae-Hwan Cho2, Hae-Kag Lee3
1Department of Radiology, Kyung Hee University Hospital at Gang-dong, Seoul, Republic of Korea ; Department of International Radiological Science, Hallym University of Graduate Studies, Seoul, Republic of Korea.
Objectives:
The study measured the dose on body regions that were not shielded to protect from radiation exposure during the general procedure, with the goal of providing basic radiation dose data for radiological technologists who perform the radiographic examination.
Materials And Methods:
Shooting parts with the phantom were similar to human tissues using general shooting equipment in the general examination room. The scattered rays were measured with the ion chamber. The hand received the highest average radiation dose and the kidney the lowest. The same pattern was evident for the average equivalent dose. The available daily shooting was highest in the anterior/posterior skull, followed by the posterior/anterior chest, abdomen, anterior/posterior spine and extremities.
Results:
The daily available numbers for the eye were lower than other body regions (6-times, 4-times, 26-times, 3-times and 121-times) and the numbers on the foot were higher than for other regions (73-times, 48-times, 263-times, 39-times and 702-times).
Conclusions:
Radiation should be thoroughly blocked by the apron to protect the radiological technologist from the radiation exposure, the proper distance from the irradiation source should be maintained exposure is inevitable and the exposure dose and working environment shall be regularly assessed to ensure minimal exposure dose of the radiological technologist in accordance with the International Commission on Radiological Protection recommendation.
Related Concept Videos
Biological Effects of Radiation
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Dose Size and Dosing Frequency: Determination Methods
X-ray Imaging