A high-throughput biliverdin assay using infrared fluorescence.
Aleš Berlec1,2, Borut Štrukelj1,2
1Department of Biotechnology, Jožef Stefan Institute, Ljubljana, Slovenia (Berlec, Štrukelj).
Summary
A new infrared fluorescent protein assay simplifies biliverdin measurement in veterinary diagnostics. This fast, inexpensive method overcomes limitations of previous assays for liver disease testing.
Area of Science:
- Biochemistry
- Analytical Chemistry
- Veterinary Medicine
Background:
- Biliverdin, a heme degradation intermediate, is crucial for diagnosing liver-related diseases in veterinary medicine.
- Current diagnostic methods for biliverdin often require complex chromatographic assays, limiting their widespread use.
- There is a need for simpler, faster, and more accessible biliverdin quantification techniques.
Purpose of the Study:
- To develop and validate a novel, high-throughput assay for biliverdin quantification.
- To utilize the interaction between biliverdin and infrared fluorescent protein (iRFP) for fluorescence-based detection.
- To establish the assay's performance characteristics, including linearity, sensitivity, accuracy, and precision.
Main Methods:
- A new assay was developed based on the interaction of biliverdin with infrared fluorescent protein (iRFP).
- The assay measures the infrared fluorescence emitted by the functional protein complex.
- Dimethyl sulfoxide (DMSO) was used for sample extraction, and interference from hemin and bilirubin was assessed.
Main Results:
- The assay demonstrated linearity in the 0-10 µmol/l biliverdin range, with a limit of detection of 0.02 µmol/l and limit of quantification of 0.03 µmol/l.
- High accuracy (relative error < 0.15) and precision (coefficient of variation < 5%) were achieved.
- Over 95% biliverdin recovery was obtained from biological samples via DMSO extraction, with minimal hemin interference.
Conclusions:
- The newly developed iRFP-based assay provides a simple, fast, and inexpensive method for biliverdin quantification.
- This assay is suitable for high-throughput analysis in veterinary clinical diagnostics.
- The assay overcomes previous limitations, enabling broader application in diagnosing liver-related diseases.