An ionic liquid-modified graphene based molecular imprinting electrochemical sensor for sensitive detection of bovine
Zonghua Wang1, Feng Li1, Jianfei Xia1
1Laboratory of Fiber Materials and Modern Textiles, the Growing Base for State Key Laboratory, Shandong Sino-Japanese Center for Collaborative Research of Carbon Nanomaterials, Collaborative Innovation Center for Marine Biomass Fiber Materials and Textiles, Qingdao University, Qingdao, Shandong 266071, China.
Abstract:
A novel kind of molecular imprinted polymers based on ionic liquid-functionalized graphene (MIPs/IL/GR) was prepared by electro-polymerization, which was applied as a molecular recognition element to modify glassy carbon electrode (GCE) to construct an electrochemical sensor (MIPs/IL/GR/GCE) for sensitive detection of bovine hemoglobin (BHb). The fabrication conditions that affect the performance of the imprinted sensor, such as pyrrole concentration, scan cycles and scan rates, have been discussed. Under the optimized conditions, the prepared molecular imprinting electrochemical sensor showed a fast rebinding dynamics, which was successfully applied to BHb detection with a wide linear range from 1.0 × 10(-10) to 1.0 × 10(-3)g/L (R=0.998) and a detection limit of 3.09 × 10(-11)g/L. Moreover, the fabricated sensor possessed a good selectivity and stability, providing a promising tool for immunoassays and clinical applications.