Application of screening experimental designs to assess chromatographic isotope effect upon isotope-coded
Szabolcs Szarka1, Katalin Prokai-Tatrai, Laszlo Prokai
1Department of Pharmacology and Neuroscience, and ‡Department of Pharmaceutical Sciences, UNT System College of Pharmacy, University of North Texas Health Science Center , 3500 Camp Bowie Boulevard, Fort Worth, Texas 76107-2699, United States.
Abstract:
Isotope effect may cause partial chromatographic separation of labeled (heavy) and unlabeled (light) isotopologue pairs. Together with a simultaneous matrix effect, this could lead to unacceptable accuracy in quantitative liquid chromatography-mass spectrometry assays, especially when electrospray ionization is used. Four biologically relevant reactive aldehydes (acrolein, malondialdehyde, 4-hydroxy-2-nonenal, and 4-oxo-2-nonenal) were derivatized with light or heavy (d3-, (13)C6-, (15)N2-, or (15)N4-labeled) 2,4-dinitrophenylhydrazine and used as model compounds to evaluate chromatographic isotope effects. For comprehensive assessment of retention time differences between light/heavy pairs under various gradient reversed-phase liquid chromatography conditions, major chromatographic parameters (stationary phase, mobile phase pH, temperature, organic solvent, and gradient slope) and different isotope labelings were addressed by multiple-factor screening using experimental designs that included both asymmetrical (Addelman) and Plackett-Burman schemes followed by statistical evaluations. Results confirmed that the most effective approach to avoid chromatographic isotope effect is the use of (15)N or (13)C labeling instead of deuterium labeling, while chromatographic parameters had no general influence. Comparison of the alternate isotope-coded derivatization assay (AIDA) using deuterium versus (15)N labeling gave unacceptable differences (>15%) upon quantifying some of the model aldehydes from biological matrixes. On the basis of our results, we recommend the modification of the AIDA protocol by replacing d3-2,4-dinitrophenylhydrazine with (15)N- or (13)C-labeled derivatizing reagent to avoid possible unfavorable consequences of chromatographic isotope effects.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
Related Concept Videos
Mass Spectrometry: Isotope Effect
Ion-Exchange Chromatography
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Types of Detectors
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...