Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Multicenter evaluation of renography with an automated physical phantom
Anssi O Nykänen1, Pentti J Rautio, Jussi V Aarnio
1aDepartment of Medical Physics, Etelä-Savo Hospital District, Mikkeli Central Hospital, Mikkeli bDepartment of Clinical Physiology, North Karelia Central Hospital, Joensuu, Finland.
Standardizing dynamic radionuclide renal imaging (renography) in Finland improved quality control. Quantitative parameter variations between laboratories decreased significantly, indicating enhanced study consistency.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Quality Control
Background:
- Dynamic radionuclide renal imaging (renography) protocols vary widely, challenging study quality.
- A need for national quality control in renography has been identified.
Purpose of the Study:
- To promote standardization of renography protocols in Finland.
- To assess improvements in renography quality since 1997.
Main Methods:
- The Heikkinen phantom was used across 20 nuclear medicine labs.
- Quantitative parameters (Tmax, T1/2, RCA20, relative uptake) were calculated and compared.
- Laboratory reports were evaluated through blind testing.
Main Results:
- Errors in Tmax and relative uptake were within acceptable ranges.
- Variations in quantitative parameters between labs decreased compared to 1997.
- Significant reductions in errors for T1/2 and RCA20 were observed.
Conclusions:
- Renography quality in Finland has improved, with reduced inter-laboratory variation.
- New software and increased quality efforts likely contributed to the observed improvements.
- Standardization efforts have successfully enhanced the reliability of renography studies.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
09:32Manufacturing Abdominal Aorta Hydrogel Tissue-Mimicking Phantoms for Ultrasound Elastography Validation
Published on: September 19, 2018
Related Concept Videos
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Imaging Studies VII: Vascular Imaging