Accuracy of amplitude-based respiratory gating for PET/CT in irregular respirations
Yuji Tsutsui1, Daisuke Kidera, Takafumi Taniguchi
1Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan, yuji1006@med.kyushu-u.ac.jp.
Objective:
We evaluated the accuracy of amplitude gating PET (AG-PET) compared with phase gating PET (PG-PET) in relation to respiratory motion patterns based on a phantom analysis.
Method:
We used a NEMA IEC body phantom filled with an (18)F solution with a 4:1 sphere-to-background radioactivity ratio (12.6 and 2.97 kBq/mL). PET/CT scans were acquired in a motionless and moving state on a Biograph mCT. The respiratory movements were simulated by four different waveform patterns consisting of ideal breathing, breathing with a pause period, breathing with a variable amplitude and breathing with a changing baseline. AG-PET selects the narrow bandwidth containing 20 % of the respiratory cycle. PG-PET was reconstructed with five gates. The image quality was physically assessed using the percent contrast (Q H,10mm), background variability (N 10mm) recovery coefficient (RC), and sphere volumes.
Result:
In regular motion patterns with ideal breathing and breathing with a pause period, the Q H,10mm, RC and sphere volumes were not different between AG-PET and PG-PET. In the variable amplitude pattern, the Q H,10mm of AG-PET was higher than that of PG-PET (35.8 vs 28.2 %), the RC of AG-PET was higher than that of PG-PET and sphere volume of AG-PET was smaller than that of PG-PET (6.4 vs 8.6 mL). In the changing baseline pattern, the Q H,10mm of AG-PET was higher than that of PG-PET (42.4 vs 16.7 %), the RC of AG-PET was higher than that of PG-PET and sphere volume of AG-PET was smaller than that of PG-PET (6.2 vs 9.8 mL). The N 10mm did not differ between AG-PET and PG-PET, irrespective of the motion pattern.
Conclusion:
Amplitude gating PET is considered to be more accurate than phase gating PET for examining unstable respiratory motion patterns, such as those involving a variable amplitude or changing baseline.
Related Concept Videos
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Respiratory Volumes and Capacities I