The impact of image reconstruction bias on PET/CT 90Y dosimetry after radioembolization
Related Experiment Video
Updated: Apr 27, 2026

Y-90 Radioembolization and PD-1 Inhibitor as Neoadjuvant Treatment in Hepatocellular Carcinoma
Published on: May 24, 2024
Katie N Tapp1, William B Lea2, Matthew S Johnson2
1Department of Radiology and Imaging Science, Indiana University School of Medicine, Indianapolis, Indiana; and School of Health Sciences, Purdue University, West Lafayette, Indiana.
Unlabelled:
PET/CT imaging after radioembolization is a viable method for determining the posttreatment (90)Y distribution in the liver. Low true-to-random coincidence ratios in (90)Y PET studies limit the quantitative accuracy of these studies when reconstruction algorithms optimized for traditional PET imaging are used. This study examined these quantitative limitations and assessed the feasibility of generating radiation dosimetry maps in liver regions with high and low (90)Y concentrations.
Methods:
(90)Y PET images were collected on a PET/CT scanner and iteratively reconstructed with the vendor-supplied reconstruction algorithm. PET studies on a Jaszczak cylindric phantom were performed to determine quantitative accuracy and minimum detectable concentration (MDC). (90)Y and (18)F point-source studies were used to investigate the possible increase in detected random coincidence events due to bremsstrahlung photons. Retrospective quantitative analyses were performed on (90)Y PET/CT images obtained after 65 right or left hepatic artery radioembolizations in 59 patients. Quantitative image errors were determined by comparing the measured image activity with the assayed (90)Y activity. PET images were converted to dose maps through convolution with voxel S values generated using MCNPX, a Monte Carlo N-particle transport code system for multiparticle and high-energy applications. Tumor and parenchyma doses and potential bias based on measurements found below the MDC were recorded.
Results:
Random coincidences were found to increase in (90)Y acquisitions, compared with (18)F acquisitions, at similar positron emission rates because of bremsstrahlung photons. Positive bias was observed in all images. Quantitative accuracy was achieved for phantom inserts above the MDC of 1 MBq/mL. The mean dose to viable tumors was 183.6 ± 156.5 Gy, with an average potential bias of 3.3 ± 6.4 Gy. The mean dose to the parenchyma was 97.1 ± 22.1 Gy, with an average potential bias of 8.9 ± 4.9 Gy.
Conclusion:
The low signal-to-noise ratio caused by low positron emission rates and high bremsstrahlung photon production resulted in a positive bias on (90)Y PET images reconstructed with conventional iterative algorithms. However, quantitative accuracy was good at high activity concentrations, such as those found in tumor volumes, allowing for adequate tumor (90)Y PET/CT dosimetry after radioembolization.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...