Computed tomography radiation dose reduction: effect of different iterative reconstruction algorithms on image quality
Related Experiment Video
Updated: Apr 27, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Martin J Willemink1, Richard A P Takx, Pim A de Jong
1From the Departments of *Radiology, and †Anatomy, University Medical Center Utrecht, Utrecht; ‡Maastricht University Medical Center, Maastricht; §Radboud University Nijmegen Medical Center, Nijmegen, The Netherlands; and ∥Department of Radiology, UZ Brussel, Brussels, Belgium.
Objective:
We evaluated the effects of hybrid and model-based iterative reconstruction (IR) algorithms from different vendors at multiple radiation dose levels on image quality of chest phantom scans.
Methods:
A chest phantom was scanned on state-of-the-art computed tomography scanners from 4 vendors at 4 dose levels (4.1 mGy, 3.0 mGy, 1.9 mGy, and 0.8 mGy). All data were reconstructed with filtered back projection (FBP) and reduced-dose data also with IR (iDose4, Adaptive Iterative Dose Reduction 3D, Adaptive Statistical Iterative Reconstruction, Sinogram-Affirmed Iterative Reconstruction, prototype Iterative Model Reconstruction, and Veo). Computed tomography numbers and noise were measured in the spine and lungs. Signal-to-noise ratios (SNR) and contrast-to-noise ratios (CNR) were calculated and differences were analyzed with the Friedman test.
Results:
For all vendors, radiation dose reduction with FBP resulted in significantly increased noise levels (≤148%) as well as decreased SNR (≤57%) and CNR (≤58%) (P < 0.001). Conversely, IR resulted in decreased noise levels (≤48%) as well as increased SNR (≤94%) and CNR (≤94%). The SNRs and CNRs of the model-based algorithms at 80% reduced dose were similar to reference-dose FBP.
Conclusions:
Hybrid IR algorithms have the potential to reduce radiation dose with 27% to 54% and model-based IR algorithms with up to 80%.
More Related Videos
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...