Measurement of electron density and effective atomic number by dual-energy scan using a 320-detector computed
Fuminari Tatsugami1, Toru Higaki, Masao Kiguchi
1From the Departments of *Diagnostic Radiology and †Radiology, Hiroshima University, Minami-ku, Hiroshima, Japan; and ‡Toshiba Medical Systems Corporation, Otawara-shi, Tochigi, Japan.
Abstract:
We evaluated the accuracy of the electron densities and effective atomic numbers determined by raw data-based dual-energy analysis on a 320-detector computed tomography scanner. The mean (SD) errors between the measured and true electron densities and between the measured and true effective atomic numbers were 1.3% (1.5%) and 3.1% (3.2%), respectively. Electron densities and effective atomic numbers can be determined with high accuracy, which may help to improve accuracy in radiotherapy treatment planning.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Imaging Studies III: Computed Tomography
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
X-ray Imaging
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET