Abdominal CT: an intra-individual comparison between virtual monochromatic spectral and polychromatic 120-kVp images
Yoshitake Yamada1, Masahiro Jinzaki1, Takahiro Hosokawa1
1Department of Diagnostic Radiology, Keio University School of Medicine, 35 Shinanomachi, Shinjuku-ku, Tokyo 160-8582, Japan.
Objectives:
To compare quantitative and subjective image quality between virtual monochromatic spectral (VMS) and conventional polychromatic 120-kVp imaging performed during the same abdominal computed tomography (CT) examination.
Materials And Methods:
Our institutional review board approved this prospective study; each participant provided written informed consent. 51 patients underwent sequential fast kVp-switching dual-energy (80/140 kVp, volume CT dose index: 12.7 mGy) and single-energy (120-kVp, 12.7 mGy) abdominal enhanced CT over an 8 cm scan length with a random acquisition order and a 4.3-s interval. VMS images with filtered back projection (VMS-FBP) and adaptive statistical iterative reconstruction (so-called hybrid IR) (VMS-ASIR) (at 70 keV), as well as 120-kVp images with FBP (120-kVp-FBP) and ASIR (120-kVp-ASIR), were generated from dual-energy and single-energy CT data, respectively. The objective image noises, signal-to-noise ratios and contrast-to-noise ratios of the liver, kidney, pancreas, spleen, portal vein and aorta, and the lesion-to-liver and lesion-to-kidney contrast-to-noise ratios were measured. Two radiologists independently and blindly assessed the subjective image quality. The results were analyzed using the paired t-test, Wilcoxon signed rank sum test and mixed-effects model with Bonferroni correction.
Results:
VMS-ASIR images were superior to 120-kVp-FBP, 120-kVp-ASIR and VMS-FBP images for all the quantitative assessments and the subjective overall image quality (all P<0.001), while VMS-FBP images were superior to 120-kVp-FBP and 120-kVp-ASIR images (all P<0.004).
Conclusions:
VMS images at 70 keV have a higher image quality than 120-kVp images, regardless of the application of hybrid IR. Hybrid IR can further improve the image quality of VMS imaging.
More Related Videos
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT