Testing quality of a self-monitoring blood glucose sensor with an auto-coding mechanism when used by patients versus
Cheng-Teng Hsu1, Ming-Hsun Wu2, Chih-Yi Kuo3
1School of Medical Laboratory and Biotechnology, Chung Shan Medical University, Taichung, Taiwan.
Background:
In response to the problem of erroneous readings due to miscoding when performing self-monitoring blood glucose (SMBG), this study introduces a user-friendly SMBG biosensor with an innovative auto-coding module on the meter and strip. Actual users characterized the performance of the SMBG systems.
Methods:
A total of 105 patients were incorporated in the study and Clarke error grid analysis (EGA) was administered to evaluate the clinical accuracy of the results obtained by the patients versus the technicians. All patients used the questionnaires to comment on the use of the auto-coding sensor.
Results:
In the imprecision test, the total CV of the 5 BG levels was 2.1%. In the EGA plot, the results of the auto-coding sensor were 96.2%, both lots A and B, in zone A for the patients and 99.0% and 97.1% for the technician. The paired t-test demonstrated no statistically significant difference between the patient and technician measurements. Regression analysis also demonstrated that the measurements taken by the patients agreed with those obtained using the laboratory method.
Conclusions:
The patients achieved satisfactory performance using the auto-coding SMBG sensor and derived similar results with both laboratory reference and operation by a technician.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Automated Microbial Diagnostics
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...