Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Quantitative evaluation of a combined pulse oximetry and end-tidal CO2 monitor
1Medical College of Wisconsin, Department of Anesthesiology, Milwaukee.
Abstract:
Interest in combined noninvasive respiratory monitors is increasing. The combination of a pulse oximeter and an end-tidal CO2 monitor in one unit reduces space requirements and, potentially, production costs. One such unit (POET, Criticare Systems, Inc.) was tested for accuracy in 15 healthy nonsmoking adult volunteers. Each patient was subjected to inspired O2 fractions of 1.0, 0.21, 0.12, 0.10, and 0.08, then to CO2 concentrations of 0, 3, 6, and 7.5%. Pulse oximetry sensors were placed on the index finger (SpfO2) and the forehead (SphO2). For each oxygen concentration tested, a 30-second steady state of SpfO2 and SphO2 was established before arterial blood was drawn for SaO2 determination. A sampling line for end-tidal carbon dioxide tension (PETCO2) was placed just distal to the one-way exhalation valve. Samples of arterial blood were drawn and analyzed for PaCO2 after PETCO2 had been stable for 30 seconds for each concentration of carbon dioxide. Linear regression was used for data analysis. The regression equations were: SpfO2 (y) vs. SaO2 (x), y = 0.93x + 6.70, r = 0.98 (p less than 0.001); SphO2 (y) vs. SaO2 (x), y = 0.85x + 13.76, r = 0.95 (p less than 0.001); PETCO2 (y) vs. PaCO2 (x), y = 1.08x - 2.66, r = 0.95 (p less than 0.001) (when subjects were not hyperventilating). The authors conclude that this combined pulse oximeter and end-tidal CO2 monitor can accurately refect SaO2 and PaCO2 in clinically useful ranges.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...