Purpose:
To estimate the value of proton resonance frequency (PRF) shift coefficient of explanted Thiel embalmed animal and human tissue used as a preclinical model for treatment with MR-guided focused ultrasound (FUS).
Methods:
Thiel embalmed human liver, ovine liver, and porcine muscle were heated using two methods: bulk heating and FUS-induced heating. Phase-referenced PRF thermometry was applied during cooling of the tissue to obtain a series of phase difference, ΔΦ, maps. A fiber-optic thermocouple was inserted in the tissue to measure the temperature difference, ΔT. The PRF shift coefficient was calculated from the measured ΔΦ, ΔT.
Results:
In the case of bulk heating, the mean values (±SD) of the PRF coefficient for Thiel embalmed ovine liver, porcine muscle, and human liver were: 0.017 (5 × 10-4 ) ppm/°C, 0.015 (6 × 10-4 ) ppm/°C, and 0.012 (6 × 10-4 ) ppm/°C, respectively. Similar values were found in tissues heated with FUS.
Conclusion:
The values of PRF coefficient measured for the Thiel embalmed tissue were higher than the values for fresh tissue, suggesting that embalming a tissue with Thiel fluid can affect PRF thermometry. The chemical composition of the Thiel fluid and the electrical conductivity may explain these results. Magn Reson Med 74:260-265, 2015. © 2014 Wiley Periodicals, Inc.
More Related Videos
07:38Real-time Monitoring of High Intensity Focused Ultrasound HIFU Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound HMIFU
Published on: November 3, 2015
05:56Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019
Related Concept Videos
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
NMR Spectroscopy: Chemical Shift Overview
For instance, the proton...
Proton (¹H) NMR: Chemical Shift
Absorption signals of all the protium nuclei...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)