Use of a Low-flow Digital Anesthesia System for Mice and Rats
Published on: September 7, 2016
Calculation of volatile anaesthetics consumption from agent concentration and fresh gas flow
1Institute of Anaesthesiology, University Hospital Zurich, Zurich, Switzerland.
Background:
The assessment of volatile agents' consumption can be performed by weighing vapourisers before and after use. This method is technically demanding and unavailable for retrospective analysis of anaesthesia records. Therefore, a method based on calculations from fresh gas flow and agent concentration is presented here.
Methods:
The presented calculation method herein enables a precise estimation of volatile agent consumption when average fresh gas flows and volatile agent concentrations are known. A pre-condition for these calculations is the knowledge of the vapour amount deriving from 1 ml fluid volatile agent. The necessary formulas for these calculations and an example for a sevoflurane anaesthesia are presented.
Results:
The amount of volatile agent vapour deriving from 1 ml of fluid agent are for halothane 229 ml, isoflurane 195 ml, sevoflurane 184 m, and desflurane 210 ml. The constant for sevoflurane is used in a fictitious clinical case to exemplify the calculation of its consumption in daily routine resulting in a total expenditure of 23.6 ml liquid agent.
Conclusions:
By application of the presented specific volatile agent constants and equations, it becomes easy to calculate volatile agent consumption if the fresh gas flows and the resulting inhaled concentration of the volatile agent are known. By this method, it is possible to extract data about volatile agent consumption both ways: (1) retrospectively from sufficiently detailed and accurate anaesthesia recordings, as well as (2) by application of this method in a prospective setting. Therefore, this method is a valuable contribution to perform pharmacoeconomical surveys.
More Related Videos
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
04:56In Vitro Method to Control Concentrations of Halogenated Gases in Cultured Alveolar Epithelial Cells
Published on: October 23, 2018
Related Concept Videos
Inhalational Anesthetics: Overview
Stages of General Anesthesia
Volatilization
One-Compartment Open Model for IV Bolus Administration: Estimation of Elimination Rate Constant, Half-Life and Volume of Distribution
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Turbulent Flow: Problem Solving
Temperature is a key factor in CO2 solubility. In this case, the CO2 gas and the liquid are cooled to 20°C. Lower temperatures...