Characterizing the reflectivity of handheld display devices
1Division of Imaging and Applied Mathematics, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Avenue, Silver Spring, Maryland 20993.
Measuring medical image display reflectivity is crucial. The American Association of Physicists in Medicine Task Group 18 (TG18) method needs adjustments for handheld devices due to illumination variability and back panel effects.
Area of Science:
- Medical Physics
- Display Technology
- Image Quality Assessment
Background:
- Increasing use of handheld and tablet devices for medical image viewing necessitates standardized reflectivity measurement methods.
- Current methods, like the American Association of Physicists in Medicine Task Group 18 (TG18), may not adequately address the unique characteristics of smaller display devices.
Purpose of the Study:
- To characterize diffuse reflections for handheld display devices (mobile phones, tablets) using modified TG18 methods.
- To assess the impact of back surface conditions and masking on diffuse reflectance coefficient measurements.
Main Methods:
- Modified the TG18 diffuse reflectance coefficient measurement protocol.
- Measured seven handheld devices and three workstation displays, varying back panel conditions and mask sizes.
- Recorded illuminance and reflected luminance to calculate diffuse reflectance coefficients.
Main Results:
- Diffuse reflectance coefficients for handhelds were lower without a back panel and increased as mask size decreased.
- Workstation displays showed higher coefficients without masks compared to masked conditions.
- Both luminance and illuminance increased with decreasing mask size, but not proportionally.
Conclusions:
- The TG18 method is sensitive to illumination conditions and back panel variations, especially for smaller handheld displays.
- Diffuse reflection coefficients can vary significantly (up to 20%) based on mask size and back surface.
- The TG18 method requires refinement, including specifying the back panel, for consistent medical display comparisons across device types.