Related Experiment Videos
Metal release from plates used in jaw fracture treatment. A pilot study
L E Moberg1, A Nordenram, O Kjellman
1Department of Prosthodontics, Karolinska Institute, Stockholm, Sweden.
Abstract:
This study investigated the occurrence of corrosion associated with the use of metallic implants to stabilize jaw fractures. Three different types of plates, Co-Cr and Ni-Cr alloys and Titanium, were connected in vivo to the mandibular bone surface of monkeys (Cercopithecus aethiops). The animals were killed after 3 and 6 months. The mucous membrane and bone tissue were analysed for concentrations of Co, Cr, Ni, Mo, Al, and Ti by atomic absorption spectrophometry and a radiochemical neutron activation technique. With the exception of Ti, higher concentrations of all the above elements were found in the tissue near the implants when compared with contralateral controls. However, no signs of corrosion, macroscopic or microscopic, were observed on the surface of the implants.