Effect of embedded dexamethasone in cochlear implant array on insertion forces in an artificial model of scala
Yann Nguyen1, Daniele Bernardeschi, Guillaume Kazmitcheff
1*INSERM, Minimally Invasive Robot-based Hearing Rehabilitation, UMR; †Otolaryngology Department, Unit of Otology, Auditory Implants and Skull Base Surgery, AP-HP, Hospital Pitie´ Salpêtrière; and ‡Sorbonne University, Université Pierre et Marie Curie, Paris, France.
Hypothesis:
Loading otoprotective drug into cochlear implant might change its mechanical properties, thus compromising atraumatic insertion. This study evaluated the effect of incorporation of dexamethasone (DXM) in the silicone of cochlear implant arrays on insertion forces.
Background:
Local administration of DXM with embedded array can potentially reduce inflammation and fibrosis after cochlear implantation procedure to improve hearing preservation and reduce long-term impedances.
Methods:
Four models of arrays have been tested: 0.5-mm distal diameter array (n = 5) used as a control, drug-free 0.4-mm distal diameter array (n = 5), 0.4-mm distal diameter array with 1% eluded DXM silicone (n = 5), and 0.4-mm distal diameter array with 10% eluded DXM silicone (n = 5). Via a motorized insertion bench, each array has been inserted into an artificial scala tympani model. The forces were recorded by a 6-axis force sensor. Each array was tested seven times for a total number of 140 insertions.
Results:
During the first 10-mm insertion, no difference between the four models was observed. From 10- to 24-mm insertion, the 0.5-mm distal diameter array presented higher insertion forces than the drug-free 0.4-mm distal diameter arrays, with or without DXM. Friction forces for drug-free 0.4-mm distal diameter array and 0.4-mm distal diameter DXM eluded arrays were similar on all insertion lengths.
Conclusion:
Incorporation of DXM in silicone for cochlear implant design does not change electrode array insertion forces. It does not raise the risk of trauma during array insertion, making it suitable for long-term in situ administration to the cochlea.