Design changes improve contact patterns and articular surface damage in total knee arthroplasty
Susannah L Gilbert1, Adam J Rana1, Joseph D Lipman1
1Hospital for Special Surgery, New York, NY, United States.
The Knee
|August 16, 2014
Summary
The Optetrak Logic knee replacement, with design modifications, significantly reduced anterior post damage compared to the original Optetrak PS. This innovation enhances the longevity of posterior stabilized knee implants.
Area of Science:
- Orthopedic surgery
- Biomedical engineering
- Materials science
Background:
- The Optetrak PS posterior stabilized knee replacement has been effective since 1995.
- In 2009, the Optetrak Logic was introduced with design modifications to reduce edge loading on the tibial post.
Purpose of the Study:
- To provide the rationale for the Optetrak Logic design change.
- To compare post damage between the Optetrak PS and Optetrak Logic designs.
Main Methods:
- Retrieval findings of tibial post damage were integrated with finite element analysis.
- A 3:1 matched sample of 60 PS and 20 Logic inserts were analyzed for polyethylene damage.
Main Results:
- Incorporating rounded contact geometries markedly reduced polyethylene stresses.
- The redesigned Optetrak Logic showed reduced surface damage and deformation on the tibial post compared to the original design.
Conclusions:
- The study demonstrates a design cycle using retrieval analysis and analytical tools to improve implant performance.
- Design changes to the post-box geometry have markedly reduced anterior post damage in posterior stabilized knee replacements.