Three-dimensional Optical-resolution Photoacoustic Microscopy
Published on: May 3, 2011
High resolution three-dimensional photoacoutic tomography with CCD-camera based ultrasound detection
Robert Nuster1, Paul Slezak2, Guenther Paltauf1
1Department of Physics, Karl-Franzens-University Graz, 8010 Graz, Austria.
Abstract:
A photoacoustic tomograph based on optical ultrasound detection is demonstrated, which is capable of high resolution real-time projection imaging and fast three-dimensional (3D) imaging. Snapshots of the pressure field outside the imaged object are taken at defined delay times after photoacoustic excitation by use of a charge coupled device (CCD) camera in combination with an optical phase contrast method. From the obtained wave patterns photoacoustic projection images are reconstructed using a back propagation Fourier domain reconstruction algorithm. Applying the inverse Radon transform to a set of projections recorded over a half rotation of the sample provides 3D photoacoustic tomography images in less than one minute with a resolution below 100 µm. The sensitivity of the device was experimentally determined to be 5.1 kPa over a projection length of 1 mm. In vivo images of the vasculature of a mouse demonstrate the potential of the developed method for biomedical applications.