Related Experiment Video
Updated: Apr 25, 2026

Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
Pulse wave analysis with two tonometric devices: a comparison study
D Agnoletti1, S C Millasseau, J Topouchian
1Paris Descartes University, AP-HP, Diagnosis and Therapeutic Center, Hôtel-Dieu, Paris, France. Department of Internal Medicine, University of Bologna, Bologna, Italy.
Abstract:
Pulse wave analysis is a pivotal tool to estimate central haemodynamic parameters. Available commercial devices use applanation tonometry and have been validated against invasive catheterism. We previously observed differences on a radial second systolic peak (rSPB2) between two commonly used devices: SphygmoCor (AtCor, Australia) and PulsePen (DiaTecne, Italy). The aim of our study was to further quantify differences in radial and carotid signals from the two devices.We measured radial and carotid waveforms in 38 patients with minimal changes between systolic, diastolic blood pressure and heart rate.rSBP2, mean pressure, form factor and augmentation index were different with SphygmoCor providing lower values (mean differences: 2.2 ± 3.8 mmHg; 1.5 ± 1.7 mmHg; 3.2 ± 3.5%; 4.2 ± 8.4%, respectively). Carotid augmentation index and form factor were similar. However carotid systolic pressure (cSBP) from PulsePen was higher than cSBP from SphygmoCor (2.7 ± 4.4 mmHg, p < 0.001). For both carotid and radial signals, harmonics moduli were similar across the spectrum with the exception of the 1st harmonic.PulsePen and SphygmoCor sensors are not equivalent and provide different wave shapes despite similar harmonics content with more discrepancy on radial derived parameters than on carotid derived parameters. Further studies are required to compare invasive pressure parameters to indices derived from these two devices.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:

