Evaluation of image quality of DWIBS versus DWI sequences in thoracic MRI at 3T
Caroline Mesmann1, Monica Sigovan2, Lise-Prune Berner2
1Department of Radiology, Centre hospitalier de la Croix-Rousse, Hospices Civils de Lyon, Lyon, France.
Purpose:
To compare diffusion weighted imaging with background suppression (DWIBS) sequence with classic spectral diffusion sequence (DWI) with and without respiratory gating in mediastinal lymph node analysis at 3T.
Materials And Methods:
26 patients scheduled for mediastinoscopic lymph node analysis, prospectively undergone a thoracic 3T MRI with DWIBS (FatSat=STIR; TR/TE=6674.1/44.7ms; IR=260 ms) and DWI sequences (FatSat=SPIR; TR/TE=1291/59.6 ms) (b=0-400-800 s/mm2) with and without (free breathing) respiratory gating. Images at b=800 were analyzed by two radiologists. They performed qualitative analysis of fat-sat homogeneity and motion artifacts, rated from 0 to 4, and quantitative evaluation by studying signal to background (STB) of lymph nodes.
Results:
Quality of fat suppression was significantly higher for DWIBS than for DWI both for free-breathing (score 3.48±0.65 vs. 1.76±0.96, p<0.0001) and respiratory-gated scans (3.17±0.77 vs. 1.72±0.73, p=0.0001). Similarly, artifacts were reduced with DWIBS (3.16±0.47 vs. 1.76±0.59, p<0.0001; 3.0±0.73 vs. 2.04±0.53, p=0.0001). Quantitative analysis showed higher STB with DWIBS (3.26±1.83 vs. 0.98±0.44, p<0.0001; 3.56±, 2.09 vs. 0.92±0.59, p<0.0001). Gating did not improve image quality and STB on DWIBS (p>0.05).
Conclusion:
In thoracic MRI, ungated DWIBS sequence improves fat-sat homogeneity, reduces motion artifacts and increases STB of lymph nodes. Respiratory gating does not improve DWIBS image quality.