Modeling of effect of glucose sensor errors on insulin dosage and glucose bolus computed by LOGIC-Insulin
Tom Van Herpe1, Bart De Moor2, Greet Van den Berghe3
1Department of Intensive Care Medicine, University Hospitals Leuven, Katholieke Universiteit Leuven, Leuven, Belgium; Department of Electrical Engineering (ESAT-SCD), iMINDS Medical Information Technologies, Katholieke Universiteit Leuven, Leuven (Heverlee), Belgium. tom.vanherpe@esat.kuleuven.be.
Background:
Effective and safe glycemic control in critically ill patients requires accurate glucose sensors and adequate insulin dosage calculators. The LOGIC-Insulin calculator for glycemic control has recently been validated in the LOGIC-1 randomized controlled trial. In this study, we aimed to determine the allowable error for intermittent and continuous glucose sensors, on the basis of the LOGIC-Insulin calculator.
Methods:
A gaussian simulation model with a varying bias (0%-20%) and CV (-20% to +20%) simulated blood glucose values from the LOGIC-1 study (n = 149 patients) in 10 Monte Carlo steps. A clinical error grid system was developed to compare the simulated LOGIC-Insulin-directed intervention with the nominal intervention (0% bias, 0% CV). The severity of error measuring the clinical effect of the simulated LOGIC-Insulin intervention was graded as type B, C, and D errors. Type D errors were classified as acutely life-threatening (0% probability preferred).
Results:
The probability of all types of errors was lower for continuous sensors compared with intermittent sensors. The maximum total error (TE), defined as the first TE introducing a type B/C/D error, was similar for both sensor types. To avoid type D errors, TEs <15.7% for intermittent sensors and <17.8% for continuous sensors were required. Mean absolute relative difference thresholds for type C errors were 7.1% for intermittent and 11.0% for continuous sensors.
Conclusions:
Continuous sensors had a lower probability for clinical errors than intermittent sensors at the same accuracy level. These simulations demonstrated the suitability of the LOGIC-Insulin control system for use with continuous, as well as intermittent, sensors.
More Related Videos
08:32Studying the Hypothalamic Insulin Signal to Peripheral Glucose Intolerance with a Continuous Drug Infusion System into the Mouse Brain
Published on: January 4, 2018
07:30Homogeneous Time-resolved Förster Resonance Energy Transfer-based Assay for Detection of Insulin Secretion
Published on: May 10, 2018
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Hypoglycemia and Glucagon
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...
Glucose Homeostasis: Pancreatic Islets and Insulin Secretion
Insulin and C-peptide are...
Hyperglycemia
Glucagon-like Receptor Agonists
GLP-1, when administered in high doses intravenously, triggers insulin secretion, inhibits glucagon release, slows gastric emptying, reduces food intake, and restores normal insulin secretion. However, its rapid inactivation by...