Differentiation of low-attenuation intracranial hemorrhage and calcification using dual-energy computed tomography in
Jessica L Nute1, Lucia Le Roux, Adam G Chandler
1From the Departments of *Imaging Physics, and †Cancer Systems Imaging, The University of Texas MD Anderson Cancer Center, Houston, TX; ‡GE Healthcare, Waukesha, WI; Departments of §Biostatistics, and ∥Diagnostic Radiology, The University of Texas MD Anderson Cancer Center, Houston, TX.
Objectives:
Calcific and hemorrhagic intracranial lesions with attenuation levels of less than 100 Hounsfield units (HUs) cannot currently be reliably differentiated by single-energy computed tomography (SECT). The proper differentiation of these lesion types would have a multitude of clinical applications. A phantom model was used to test the ability of dual-energy CT (DECT) to differentiate such lesions.
Materials And Methods:
Agar gel-bound ferric oxide and hydroxyapatite were used to model hemorrhage and calcification, respectively. Gel models were scanned using SECT and DECT and organized into SECT attenuation-matched pairs at 16 attenuation levels between 0 and 100 HU. Dual-energy CT data were analyzed using 3-dimensional (3D) Gaussian mixture models (GMMs), as well as a simplified threshold plane metric derived from the 3D GMM, to assign voxels to hemorrhagic or calcific categories. Accuracy was calculated by comparing predicted voxel assignments with actual voxel identities.
Results:
We measured 6032 voxels from each gel model, for a total of 193,024 data points (16 matched model pairs). Both the 3D GMM and its more clinically implementable threshold plane derivative yielded similar results, with higher than 90% accuracy at matched SECT attenuation levels of 50 HU and greater.
Conclusions:
Hemorrhagic and calcific lesions with attenuation levels between 50 and 100 HU were differentiable using DECT in a clinically relevant phantom system with higher than 90% accuracy. This method warrants further testing for potential clinical applications.