Absorbed radiation dose in radiosensitive organs using 64- and 320-row multidetector computed tomography: a
Atif N Khan1, Waqas Shuaib2, Boris Nikolic3
1Department of Radiology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA 02215, USA.
The 320-row detector mode significantly reduces absorbed radiation dose (ARD) in radiosensitive organs compared to the 64-row mode during ECG-gated MDCTA scans. This dose reduction is observed in both prospective and retrospective gating protocols.
Area of Science:
- Medical Imaging
- Radiological Physics
- Radiation Dosimetry
Background:
- Minimizing radiation exposure is crucial in medical imaging, especially for radiosensitive organs.
- ECG-gated MDCTA scanners offer dose modulation techniques to reduce patient radiation dose.
- Detector row configurations (e.g., 64-row vs. 320-row) can impact radiation dose distribution.
Purpose of the Study:
- To compare the absorbed radiation dose (ARD) in radiosensitive organs using prospective and full-phase dose modulation techniques.
- To evaluate the impact of 64-row versus 320-row detector modes on ARD during ECG-gated MDCTA.
- To determine optimal scanning parameters for dose reduction.
Main Methods:
- A female phantom was utilized to measure organ radiation dose.
- Five DP-3 radiation detectors measured ARD in the lungs, breast, and thyroid.
- The Aquilion ONE scanner was used in 64- and 320-row modes with prospective and retrospective gating.
- Measurements were conducted at varying tube voltages (100, 120, 135 kVp), current (400 mA), and heart rates (60, 75 bpm).
Main Results:
- The 320-row mode consistently showed lower ARD than the 64-row mode across all imaging protocols.
- Prospective gating with the 320-row mode yielded significantly lower ARD for mid-breast, lung, and thyroid compared to the 64-row mode (P < 0.001).
- Retrospective gating also demonstrated lower ARD with the 320-row mode compared to the 64-row mode (P < 0.001).
- Lower kVp and heart rate settings further reduced ARD.
Conclusions:
- The 320-row detector mode offers superior dose reduction for radiosensitive organs compared to the 64-row mode in ECG-gated MDCTA.
- Both prospective and retrospective gating protocols benefit from the 320-row configuration for dose optimization.
- The 64-row mode's performance was comparable to other 64-row MDCT scanners.
More Related Videos
06:52Positron Emission Tomography Using 64-Copper as a Tracer for the Study of Copper-Related Disorders
Published on: April 28, 2023
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
X-ray Imaging