Cleanliness of disposable vs nondisposable electrocardiography lead wires in children

Nancy Addison1, Beth Quatrara2, Lisa Letzkus2

  • 1Nancy Addison is a research nurse clinician in the pediatric intensive care unit, Beth Quatrara is director and Lisa Letzkus is assistant director of nursing research for the medical center, David Strider is a nurse practitioner at the medical center, Virginia Rovnyak is a statistician at the school of nursing, and Virginia Syptak and Lisa Fuzy are nurse clinicians in the pediatric intensive care unit. All are employed at the University of Virginia Health System in Charlottesville. nkg2x@hscmail.mcc.virginia.edu.

Insights

Disposable electrocardiography (ECG) lead wires show lower adenosine triphosphate (ATP) levels in pediatric cardiac surgery patients compared to nondisposable wires. This suggests disposable ECG leads can improve hospital cleanliness and reduce infection risks.

Area of Science:

  • Medical Device Cleanliness
  • Hospital Infection Control
  • Pediatric Cardiac Surgery

Background:

  • Mediastinitis poses significant financial and clinical burdens on hospitals.
  • Limited data exists on adenosine triphosphate (ATP) levels on electrocardiography (ECG) lead wires in pediatric patients.
  • ATP measurement is a standard for assessing hospital equipment cleanliness and microbial contamination.

Purpose of the Study:

  • To compare the cleanliness of disposable versus nondisposable ECG lead wires.
  • To quantify adenosine triphosphate (ATP) levels on lead wires in pediatric cardiac surgery patients.
  • To evaluate ATP as a marker for microbial contamination on medical devices.

Main Methods:

  • Prospective, randomized trial design.
  • ATP swab testing on ECG lead wires (disposable and nondisposable).
  • Testing conducted on postoperative days 1, 2, and 3 in 51 pediatric patients.

Main Results:

  • Disposable ECG lead wires demonstrated significantly lower ATP counts on postoperative day 1 (median 157 RLUs vs. 610 RLUs).
  • Lower ATP levels were observed for disposable wires on postoperative day 2 (median 200 RLUs vs. 453 RLUs).
  • Results indicate a statistically significant difference in cleanliness between the two types of lead wires.

Conclusions:

  • Disposable ECG lead wires are a cleaner option for pediatric cardiac surgery patients.
  • Utilizing disposable ECG lead wires can reduce ATP contamination for at least 48 hours post-surgery.
  • This strategy may contribute to improved patient safety and reduced hospital-acquired infections.
Abstract