Technical note: a method for improving the calibration reproducibility of an ionization chamber detector array
Related Experiment Video
Updated: Apr 24, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
1Brighton and Sussex University Hospitals NHS Trust, Eastern Road, Brighton BN2 5BE, United Kingdom.
Purpose:
This paper describes an extension to a wide field calibration method implemented on a commercial detector array in order to improve the reproducibility of the calibration procedure.
Methods:
Following the standard array calibration procedure, two additional 10×10 cm exposures were acquired for each array axis with the detector array shifted by ±10 cm in the transverse or axial axes, or by ±10√2 cm in the positive or negative diagonal axes. These exposures were compared with a final baseline 10×10 cm exposure captured with the detector repositioned at the isocenter. The measurements were used to calculate a linear off-axis correction gradient which was then applied to the stored calibration factors.
Results:
The mean coefficient of variation between five repeat calibrations was reduced from 4.17% to 0.48% and the maximum percentage error in individual calibration factors was reduced from 6.46% to 0.77%.
Conclusions:
The reproducibility of the calibration factors of an ionization chamber array was increased by capturing a baseline exposure and two further off-axis readings per calibration axis.
Related Concept Videos
Mass Analyzers: Common Types
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...

