Curtain Flow Column: Optimization of Efficiency and Sensitivity
Published on: June 12, 2016
Minimizing ultraviolet noise due to mis-matches between detector flow cell and post column mobile phase temperatures
1SFC Solutions, Inc., 9435 Downing St, Englewood, FL 34224, United States.
Abstract:
A mis-match between the post-column mobile phase temperature and the UV detector flow cell temperature can cause significant UV noise in supercritical fluid chromatography (SFC). Deviations as little as 5 °C can increase noise as much as 5 times, making the detector unsuited for trace analysis. Two approaches were used to minimize this noise. When a flow cell was in direct thermal contact (metal on metal) with the detector optical bench, the mobile phase temperature was actively controlled to the measured flow cell temperature, by using one of the heat exchangers (HX) in the column compartment. However, with some older, but still widely used flow cell designs, this required repeated, hourly monitoring of the flow cell temperature and repeated manual adjustment of the heat exchanger temperature, due to thermal drift. Flow cell design had a strong influence on susceptibility to this thermally induced noise. Thermally insulating the flow cell from the optical bench made some cells much less susceptible to such thermally induced noise. Five different flow cells, some insulated, some un-insulated, were evaluated. Most had a truncated conical flow path, but one had a cylindrical flow path. Using either approach, the ASTM noise, with a 10mm, 13 μL conical flow cell, could be optimized to ≈0.007 mAU at 2.5 Hz, in SFC, which is very near the 0.006 mAU manufacturer's specification for HPLC. The insulated version of this flow cell required far less optimization, compared to the un-insulated version. At 150 bar, an experimental 3mm, 2 μL flow cell, with only one side insulated, yielded noise slightly too high (≈0.16-0.18 mAU) for trace analysis, at 80 Hz. However, at 200 bar, noise at 80 Hz was <0.06 mAU, which should allow quantification of a 1 mAU tall trace component with a signal to noise ratio (S/N) >10. Even partially un-insulated, this flow cell design was much less susceptible to thermally induced noise. Further insulating this flow cell design failed to improve performance.
Related Concept Videos
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: