Determination of volatile organic and polycyclic aromatic hydrocarbons in crude oil with efficient
Haijing Wang1, Helmut Geppert2, Thomas Fischer2
1Central Analytical Laboratory, Brandenburg University of Technology Cottbus-Senftenberg, Konrad-Wachsmann-Allee 6, Cottbus 03046, Germany whj2011@gmail.com.
Abstract:
Determination of volatile organic compounds (VOCs) in crude oil, such as super volatile organic compounds (super VOCs) and simple polycyclic aromatic hydrocarbons (PAHs), is vital for targeting crude oil spill spots. In this study, a static headspace gas chromatography flame ionization detection method was established for determination of super VOCs in crude oil with both external and internal standard determination, which can be used in the field when using portable gas chromatography. Identification was done by comparing the retention time with the corresponding standards and quantitation was done with a new one-drop method. Another simplified and efficient method was performed to analyze volatile PAHs in crude oil, which can also be used in field analysis. Toluene was used as the extraction solvent for PAHs in crude oil. Method validation for both analyses was satisfactory. The result showed that n-butane and n-pentane were maximum super VOCs and naphthalene, phenanthrene and fluorene were the main PAHs in the crude oil studied. The super VOCs quantity ranged from 3 to 6% and the main PAHs consisted of 0.02-0.06% of studied crude oil.
More Related Videos
11:44Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
Published on: March 6, 2016
07:49On-line Analysis of Nitrogen Containing Compounds in Complex Hydrocarbon Matrixes
Published on: August 5, 2016
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Volatilization
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Aromatic Compounds: Overview
In 1825, Faraday...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...