Related Experiment Video
Updated: Apr 23, 2026

Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
Quantification of myocardial blood flow using non-ECG-triggered MR imaging
David Chen1,2, Behzad Sharif2, Rohan Dharmakumar2,3
1Department of Biomedical Engineering, Northwestern University, Chicago, Illinois, USA.
Purpose:
MR myocardial perfusion imaging is dependent on reliable electrocardiogram (ECG) triggering for accurate measurement of myocardial blood flow (MBF). A non-ECG-triggered method for quantitative first-pass imaging may improve clinical feasibility in patients with poor ECG signal. The purpose of this study is to evaluate the feasibility of a non-ECG-triggered method for myocardial perfusion imaging in a single slice.
Methods:
The proposed non-ECG-triggered technique uses a saturation-recovery magnetization preparation and golden-angle radial acquisition for integrated arterial input function (AIF) measurement. Image based self-gating with a temporal resolution of 42.6 ms is used to generate a first-pass image series with consistent cardiac phase. The AIF is measured using beat-by-beat T1 estimation of the ventricular blood pool. The proposed technique was performed on 14 healthy volunteers and compared against a conventional ECG-triggered dual-bolus acquisition.
Results:
The proposed method produced MBF with no significant difference compared with ECG-triggered technique (mean of 0.63 ± 0.22 mL/min/g to 0.73 ± 0.21 mL/min/g).
Conclusion:
We have developed a non-ECG-triggered perfusion imaging method with T1 based measurement of the AIF in a single slice. In this preliminary study, our results demonstrate that MBF measured using the proposed method is comparable to the conventional ECG-triggered method.
More Related Videos
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
06:56Human Fetal Blood Flow Quantification with Magnetic Resonance Imaging and Motion Compensation
Published on: January 7, 2021
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI