Assessment of effective face mask ventilation is compromised during synchronised chest compressions
Trang Huynh1, Rae Jean Hemway2, Jeffrey M Perlman1
1Division of Newborn Medicine, Department of Pediatrics, New York Presbyterian Hospital, New York, USA.
Background:
Delivery room cardiopulmonary resuscitation is rare. Recent evidence suggests that effective ventilation may be compromised during chest compressions (CC).
Objectives:
To determine whether trained neonatal personnel can assess effective ventilation during CC in the setting of changing lung compliance.
Methods:
Neonatal providers (n=30) provided CC using a 3:1 CC to ventilation ratio performed for 2 min, with lung compliance adjusted every 30 s from 0.5 (low) to 1.0 mL/cmH2O (normal), followed by face mask ventilation (FMV) alone for 1 min. A neonatal lung simulator connected to a neonatal manikin was used to simulate the volume/pressure relation at low and normal compliance.
Results:
Group analysis showed no difference in peak inflating pressure (PIP) at low versus normal compliance, but a threefold increase in tidal volume (TV) (p=0.00005) during synchronised CC. Paired analysis demonstrated minimal change in PIP, but a significant decrease in TV at low versus normal compliance. During FMV only, a significant decrease in PIP and increase in TV was noted with improved compliance. The face mask was incorrectly applied in 12 (40%) cases and in 20/30 (67%) providers did not perceive a change in compliance. During FMV only, 7/30 (23%) took corrective steps to achieve chest rise.
Discussion:
Most providers cannot assess the effectiveness of delivered TV in the face of changing compliance during synchronised CC, limiting the ability to make appropriate and necessary adjustments. This may prolong cardiopulmonary resuscitation and result in escalating therapies unrelated to the delivery of effective ventilation.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...