Related Experiment Video
Updated: Apr 23, 2026

Fabrication of Carbon Nanotube High-Frequency Nanoelectronic Biosensor for Sensing in High Ionic Strength Solutions
Published on: July 22, 2013
Electrochemical determination of Sudan I in food products using a carbon nanotube-ionic liquid composite modified
Benzhi Liu1, Chuntao Yin, Min Wang
1a School of Environmental Science and Engineering , Yancheng Institute of Technology , Yancheng , Jiangsu Province , China.
Abstract:
A sensitive and convenient electrochemical method was developed for the determination of Sudan I using a carbon nanotube-ionic liquid composite modified electrode with the enhancement effect of cetyltrimethyl ammonium bromide (CTAB). The modified electrode exhibited an obvious electrocatalytic activity towards the oxidation of Sudan I, and the oxidation peak current significantly increased in the presence of CTAB. The experimental parameters, such as solution pH, concentration of CTAB and accumulation time, were optimised for Sudan I determination. The oxidation peak current showed a linear relationship with the concentration of Sudan I in the range of 3.0 × 10(-8) to 3.1 × 10(-6) mol l(-1), with a detection limit of 8.0 × 10(-9) mol l(-1). The proposed method was successfully applied for the determination of Sudan I in food products of ketchup and chilli sauce.