Scan time reduction in ²³Na-Magnetic Resonance Imaging using the chemical shift imaging sequence: Evaluation of an
Sebastian Weingärtner1, Friedrich Wetterling2, Simon Konstandin1
1Computer Assisted Clinical Medicine, University Medical Center Mannheim, Heidelberg University, Mannheim, Germany.
Aim:
To evaluate potential scan time reduction in (23)Na-Magnetic Resonance Imaging with the chemical shift imaging sequence (CSI) using undersampled data of high-quality datasets, reconstructed with an iterative constrained reconstruction, compared to reduced resolution or reduced signal-to-noise ratio.
Materials And Methods:
CSI (23)Na-images were retrospectively undersampled and reconstructed with a constrained reconstruction scheme. The results were compared to conventional methods of scan time reduction. The constrained reconstruction scheme used a phase constraint and a finite object support, which was extracted from a spatially registered (1)H-image acquired with a double-tuned coil. The methods were evaluated using numerical simulations, phantom images and in-vivo images of a healthy volunteer and a patient who suffered from cerebral ischemic stroke.
Results:
The constrained reconstruction scheme showed improved image quality compared to a decreased number of averages, images with decreased resolution or circular undersampling with weighted averaging for any undersampling factor. Brain images of a stroke patient, which were reconstructed from three-fold undersampled k-space data, resulted in only minor differences from the original image (normalized root means square error < 12%) and an almost identical delineation of the stroke region (mismatch < 6%).
Conclusion:
The acquisition of undersampled (23)Na-CSI images enables up to three-fold scan time reduction with improved image quality compared to conventional methods of scan time saving.
Related Concept Videos
NMR Spectroscopy: Chemical Shift Overview
For instance, the proton...
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
NMR Spectrometers: Resolution and Error Correction
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Magnetic Resonance Imaging
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...