Influence of dynamic leaks in volume-targeted pressure support noninvasive ventilation: a bench study
Manel Luján1, Ana Sogo2, Carles Grimau2
1Servei de Pneumologia, Corporaciò Sanitària Parc Taulí, Hospital de Sabadell, Universitat Autònoma de Barcelona, Parc Taulí, Sabadell Centro de Investigación Biomédica en Red Enfermedades Respiratorias, Instituto de Salud Carlos III, Madrid mlujan@tauli.cat.
Introduction:
The effect of leaks on volume-targeted pressure support noninvasive ventilation mode has only been tested with continuous simulated leaks. The objective of the study was to assess the influence of random leaks occurring either during inspiration or expiration.
Methods:
Analysis of the volume-targeted pressure support mode in 6 commercial ventilators with single-limb circuits and intentional leak in a bench study (restrictive model). Unintentional leaks were introduced through a mechanical system during inspiration (threshold valve with 2 levels of leaks) or during expiration (active valve). Results of delivered tidal volume (VT) and pressure support were externally recorded. A pre-set VT of 550 mL was programmed, with a wide range of pressure support values.
Results:
All the ventilators showed a deviation of delivered versus programmed VT below 10% in the period without unintentional leaks. In the model with unintentional inspiratory leaks, a progressive drop in delivered VT and pressure support was observed for all ventilators. The reduction in the delivered VT for the highest inspiratory leak ranged between 21 and 40%, corresponding to a decrease in pressure support between 3.09 and 10.15 cm H2O after 5 min. Conversely, in the expiratory model, increases in delivered VT and pressure support were observed, ranging between 16 and 33% and between 2.7 and 6.5 cm H2O, respectively.
Conclusions:
The introduction of random leaks influences the performance of commercial ventilators with single-limb circuits and intentional leak. The decrease in delivered VT with inspiratory leaks reaches a magnitude that may have clinically important impacts.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...