Patient-ventilator asynchronies beyond the waveform
Candelaria de Haro1,2, Eduardo Mireles-Cabodevila3, Gerard Castells-Tortosa1,2
1Critical Care Department, Parc Taulí Hospital Universitari, Institut d'Investigació i Innovació Parc Taulí (I3PT-CERCA), Universitat Autònoma de Barcelona, Sabadell.
Purpose Of Review:
Patient-ventilator asynchrony is commonly treated as a technical mismatch requiring correction, although asynchronous breaths are not uniformly harmful and apparent synchrony may conceal injurious effort. This review summarizes evidence from the past 18 months that redefines asynchrony through inspiratory effort, the transition to assisted breathing, and continuous waveform monitoring.
Recent Findings:
Potential harm depends on effort magnitude and timing, breath stacking, ventilatory mode, lung and diaphragm vulnerability, dyspnea, and cumulative exposure. Double triggering, reverse triggering, and ineffective efforts may indicate injury, overassistance, or recovery depending on context. During transition, work shifts from ventilator to patient, exposing patients to both excessive and insufficient effort. Continuous analysis can identify burden, clustering, and physiological trajectories missed by intermittent inspection. Artificial-intelligence models increasingly estimate effort and dynamic driving pressure, but heterogeneous definitions, limited external validation, artifacts, and uncertain actionability constrain implementation.
Summary:
Management should move beyond eliminating individual events toward minimizing time outside individualized ranges of acceptable effort and assistance. Monitoring should integrate waveforms with effort, lung stress, diaphragm loading, dyspnea, ventilator settings, and temporal context. Artificial intelligence should support bedside reassessment rather than autonomous adjustment. Multicenter validation and physiology-guided trials must establish whether this approach improves patient-centered outcomes.