Multilayer coated grazing incidence condenser for large numerical aperture objective at wavelength of 4.5 nm
Abstract:
A grazing incidence condenser is developed for objectives with large numerical aperture working in Carbon-window wavelength region (λ=4.4-5.0 nm) with the use of a point light source. The condenser is composed of four pieces of toroidal mirrors and a piece of the mirror was fabricated to evaluate the performance of the mirror. The radii of the toroidal mirror are determined by ray-trace calculation, and each radius of the mirror substrate and the roughness of the polished surface were evaluated to satisfy the designed parameter. A Co/C reflection multilayer is also designed to reflect soft x-ray light at 4.5 nm wavelength, and the reflection multilayer was deposited on the mirror surface. Measured reflectance of the toroidal mirror with the reflection multilayer is higher than 0.32 at 4.5 nm wavelength.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...