Thermal clearance blood flow sensor--sensitivity, linearity and flow depth discrimination.

S Thalayasingam, D T Delpy

    Medical & Biological Engineering & Computing
    |July 1, 1989
    PubMed
    Summary

    A novel thermal clearance blood flow sensor measures flow at four depths. Laboratory tests confirm its linear response across physiological ranges, though depth sensitivity is limited by materials.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Guest editorial: special section on near infrared spectroscopy and imaging of tissues.

    Journal of biomedical optics·2012
    Same author

    Recent advances in fetal near infrared spectroscopy.

    Journal of biomedical optics·2012
    Same author

    Experimental validation of Monte Carlo and finite-element methods for the estimation of the optical path length in inhomogeneous tissue.

    Applied optics·2010
    Same author

    Time-resolved optical imaging of a solid tissue-equivalent phantom.

    Applied optics·2010
    Same author

    Monte Carlo simulations of coherent backscatter for identification of the optical coefficients of biological tissues in vivo.

    Applied optics·2010
    Same author

    Spectral dependence of temporal point spread functions in human tissues.

    Applied optics·2010

    Area of Science:

    • Biomedical Engineering
    • Physiological Measurement
    • Sensor Technology

    Background:

    • Accurate measurement of skin blood flow is crucial for understanding physiological processes and diagnosing conditions.
    • Existing methods may lack the ability to differentiate blood flow at various depths within the skin.
    • Thermal clearance techniques offer a non-invasive approach to assess blood flow.

    Purpose of the Study:

    • To design and evaluate a novel thermal clearance sensor capable of measuring blood flow at multiple depths.
    • To assess the sensor's static and dynamic performance characteristics.
    • To investigate the factors influencing the sensor's depth sensitivity.

    Main Methods:

    • Development of a thermal clearance sensor with four distinct sensing elements.

    Related Experiment Videos

  • Conducting static and dynamic laboratory tests to evaluate sensor response.
  • Analyzing the relationship between sensor geometry, material properties, and depth sensitivity.
  • Main Results:

    • The designed sensor successfully measured blood flow changes at four different depths.
    • Laboratory tests demonstrated a linear response to blood flow within the physiological range.
    • Sensor geometry significantly impacts depth response, with conventional materials limiting sensitivity to 0.2 mm.

    Conclusions:

    • The developed multi-depth thermal clearance sensor shows promise for physiological skin blood flow assessment.
    • The study highlights the importance of sensor geometry in achieving depth-specific measurements.
    • Material limitations currently restrict the achievable depth sensitivity of this sensor technology.