Evaluation of a novel brachial cuff-based oscillometric method for estimating central systolic pressure in

Pantelis A Sarafidis1, Panagiotis I Georgianos, Antonios Karpetas

  • 1Department of Nephrology, Hippokration Hospital, Aristotle University of Thessaloniki, Thessaloniki, Greece.

Insights

The Mobil-O-Graph device shows good agreement for aortic blood pressure and augmentation index in hemodialysis patients, but slightly underestimates pulse wave velocity compared to the Sphygmocor device.

Area of Science:

  • Cardiovascular Physiology
  • Nephrology
  • Medical Devices

Background:

  • Elevated wave reflections, arterial stiffness, and ambulatory blood pressure (BP) are key cardiovascular risk factors in end-stage renal disease (ESRD).
  • Accurate measurement of these parameters is crucial for risk stratification in ESRD patients undergoing hemodialysis.
  • The Mobil-O-Graph is a novel oscillometric device for assessing aortic BP, augmentation index (AIx), and pulse wave velocity (PWV).

Purpose of the Study:

  • To validate the Mobil-O-Graph device against a gold-standard tonometric device (Sphygmocor) in hemodialysis patients.
  • To assess the agreement and correlation of aortic BP, AIx, and PWV measurements between the two devices.

Main Methods:

  • Seventy-three hemodialysis patients participated in the study.
  • Measurements of aortic SBP (aSBP), heart rate-adjusted AIx (AIx(75)), and PWV were taken using both Mobil-O-Graph and Sphygmocor devices in a randomized order.
  • Sphygmocor calibration utilized brachial BP measured by a mercury sphygmomanometer.

Main Results:

  • No significant differences were found between Mobil-O-Graph and Sphygmocor for aSBP (p=0.068) and AIx(75) (p=0.229).
  • A slight underestimation of PWV was observed with Mobil-O-Graph compared to Sphygmocor (p<0.01).
  • High correlations were found for all parameters: aSBP (r=0.770), AIx(75) (r=0.400), and PWV (r=0.739). Bland-Altman analysis indicated acceptable agreement for aSBP and AIx(75), with no systemic bias for PWV.

Conclusions:

  • The Mobil-O-Graph demonstrates acceptable agreement for measuring aSBP and AIx(75) in hemodialysis patients, similar to other populations.
  • While PWV is slightly underestimated by the Mobil-O-Graph, the device shows potential for assessing arterial hemodynamics in this patient group.
Abstract

Related Concept Videos

Sites for measuring blood pressure01:21

Sites for measuring blood pressure

Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
3.8K
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
2.0K
Assessment of blood pressure in brachial artery(one-step method)01:15

Assessment of blood pressure in brachial artery(one-step method)

This procedural guide systematically measures blood pressure using an oscillometric digital sphygmomanometer, emphasizing accuracy, patient safety, and comfort.
Prepare for the Procedure:
1.3K
Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
2.9K
Measurement of Blood Pressure01:17

Measurement of Blood Pressure

Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
3.9K