Some causes of inter-laboratory variation in the results of comet assay
Nikolai P Sirota1, Aliy K Zhanataev2, Elena A Kuznetsova1
1Institute of Theoretical and Experimental Biophysics of Russian Academy of Sciences, Pushchino, Russia.
Abstract:
We performed an inter-laboratory study to determine the variation of comet assay results and to identify its possible reasons. An exchange of slides between Labs in different stages of the comet assay protocol was performed. Because identical slides, durations of alkali treatment and electrophoresis, and similar electric field strengths (2.0 V/cm and 2.14 V/cm) were used, we concluded that the observed inter-laboratory difference in the results is directly associated with the electrophoresis step. In Lab 1, mouse bone marrow cells were exposed to methyl methanesulfonate at concentrations of 10, 25 and 50 μM for 3 h at 37 °C. In Lab 2, cells the same as in Lab 1 were immobilized in LMA on slides and exposed to X-rays at doses of 3-8 Gy. We found that the transportation of slides after lysis or electrophoresis step, as well as different dyes used for scoring did not produce any significant effect on the results. No substantial difference in the data was also revealed when various software packages were used for image analysis. The temperature of the alkaline solution was shown to increase during electrophoresis and, besides, the temperature heterogeneity of the solution took place in the area of the platform, with a maximum in the middle of the chamber. The temperature heterogeneity could affect the rate of conversion of alkali labile sites into single stranded breaks. Thus, it was clearly indicated that real temperature variations during the alkali treatment and electrophoresis were an essential factor in the variability of the results between our Labs.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Propagation of Uncertainty from Systematic Error