A novel high-temperature combustion based system for stable isotope analysis of dissolved organic carbon in aqueous
F M S A Kirkels1, C Cerli, E Federherr
1University of Amsterdam, Institute for Biodiversity and Ecosystem Dynamics, Earth Surface Science, Science Park 904, 1098 XH, Amsterdam, The Netherlands.
Rationale:
Dissolved organic carbon (DOC) plays an important role in carbon cycling, making precise and routine measurement of δ(13)C values and DOC concentration highly desirable. A new promising system has been developed for this purpose. However, broad-scale application of this new technique requires an in-depth assessment of analytical performance, and this is described here.
Methods:
A high-temperature combustion Total Organic Carbon analyzer was interfaced with continuous flow isotope ratio mass spectrometry (TOC/IRMS) for the simultaneous analysis of the bulk DOC concentration and δ(13)C signature. The analytical performance (precision, memory effects, linearity, volume/concentration effects, accuracy) was thoroughly evaluated, including realistic and challenging conditions such as low DOC concentrations and natural DOC.
Results:
High precision (standard deviation, SD predominantly ≤ 0.15 ‰) and accuracy (R(2) = 0.9997) were achieved for the δ(13)C analysis of a broad diversity of DOC solutions. Simultaneously, good results were obtained for the measurement of DOC concentration. Assessment of natural abundance and slightly (13)C-enriched DOC, a wide range of concentrations (~0.2-150 mgC/L) and injection volumes (0.05-3 mL), demonstrated minor/negligible memory effects, good linearity and flexible usage. Finally, TOC/IRMS was successfully applied to determine low DOC concentrations (<2 mgC/L) and DOC from diverse terrestrial, freshwater and marine environments (SD ≤ 0.23 ‰).
Conclusions:
TOC/IRMS enables fast and reliable measurement of DOC concentrations and δ(13)C values in aqueous samples, without pre-concentration and freeze-drying. Further investigations should focus on complex, saline matrices and very low DOC concentrations, to achieve a potential lower limit of 0.2 mgC/L. Thus, TOC/IRMS will give DOC research in terrestrial and aquatic environments a huge impulse with high-resolution, routine δ(13)C analysis.
More Related Videos
11:44Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
Published on: March 6, 2016
07:49On-line Analysis of Nitrogen Containing Compounds in Complex Hydrocarbon Matrixes
Published on: August 5, 2016
Related Concept Videos
Sample Preparation for Analysis: Advanced Techniques
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Flame Photometry: Lab