A novel high-temperature combustion interface for compound-specific stable isotope analysis of carbon and nitrogen
E Federherr1,2, S Willach2, N Roos3
1Research and Innovation, Elementar Analysensysteme GmbH, Donaustr. 7, 63452, Hanau, Germany.
Rationale:
In aqueous samples compound-specific stable isotope analysis (CSIA) plays an important role. No direct method (without sample preparation) for stable nitrogen isotope analysis (δ(15) N SIA) of non-volatile compounds is known yet. The development of a novel HPLC/IRMS interface based on high-temperature combustion (HTC) for both δ(13) C and δ(15) N CSIA and its proof of principle are described in this study.
Methods:
To hyphenate high-performance liquid chromatography (HPLC) with isotope ratio mass spectrometry (IRMS) a modified high-temperature combustion total organic carbon analyzer (HTC TOC) was used. A system to handle a continuously large amount of water (three-step drying system), favorable carrier and reaction gas mix and flow, an efficient high-temperature-based oxidation and subsequent reduction system and a collimated beam transfer system were the main requirements to achieve the necessary performance.
Results:
The proof of principle with caffeine solutions of the system succeeded. In this initial testing, both δ(13) C and δ(15) N values of tested compounds were determined with precision and trueness of ≤0.5 ‰. Further tests resulted in lower working limit values of 3.5 μgC for δ(13) C SIA and 20 μgN for δ(15) N SIA, considering an accuracy of ±0.5 ‰ as acceptable.
Conclusions:
The development of a novel HPLC/IRMS interface resulted in the first system reported to be suitable for both δ(13) C and δ(15) N direct CSIA of non-volatile compounds. This highly efficient system will probably open up new possibilities in SIA-based research fields. Copyright © 2016 John Wiley & Sons, Ltd.
Related Concept Videos
Mass Spectrometry: Isotope Effect
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
High-Resolution Mass Spectrometry (HRMS)
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Mass Spectrum: Interpretation
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences