Use of 3-nitrobenzonitrile as an additive for improved sensitivity in sonic-spray ionization mass spectrometry
Katerina Kanaki1, Spiros A Pergantis
1Environmental Chemical Processes Laboratory, Department of Chemistry, University of Crete, Voutes Campus, Heraklion, 71003, Greece.
Rationale:
Sonic-spray ionization (SSI) has been shown to produce gas-phase ions for a wide range of compounds, without the application of voltage or a laser. However, it remains to be shown that it can also provide similar sensitivities to those obtained by electrospray ionization mass spectrometry (ESI-MS).
Methods:
Here we report on an attempt to further improve the sensitivity of SSI-MS, more specifically a version of SSI that is referred to as Venturi easy ambient sonic-spray ionization (V-EASI) MS, by adding a signal-enhancing additive to the sample solution. The additive used is 3-nitrobenzonitrile (3-NBN), which has recently been used with success in a new ionization approach named matrix-assisted ionization vacuum. In order to conduct this study we have analyzed a range of compounds, including peptides, metalloproteins, and some organometalloids. During the V-EASI-MS analyses molecular ion and protonated molecule signal intensities as well as their corresponding signal-to-noise (S/N) ratios, obtained in the presence and absence of the 3-NBN, were compared.
Results:
The 3-NBN-assisted V-EASI-MS approach developed here provides significant improvement in sensitivity relative to conventional V-EASI-MS for almost all compounds tested. More specifically, for peptides a 1.6- to 4-fold enhancement was realized, for proteins the enhancements were from 2- to 5-fold, and for some metalloid species enhancements reached up to 10-fold. However, optimum additive concentration and ion transfer capillary temperature were found to be compound-dependent and thus require optimization in order for maximum enhancements to be achieved. In most cases the 3-NBN-assisted V-EASI-MS approach provides comparable sensitivities and S/N ratios to ESI-MS on the same ion trap mass spectrometer.
Conclusions:
The use of 3-NBN with V-EASI-MS gives rise to a novel 3-NBN-assisted MS technique, which has demonstrated considerable signal enhancement for most of the compounds analyzed, thus improving its competitiveness towards the well-established and dominating ESI-MS technique.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Chemical Ionization (CI) Mass Spectrometry
Electrophilic Aromatic Substitution: Nitration of Benzene
2° Amines to N-Nitrosamines: Reaction with NaNO2
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences