1Department of Nuclear Medicine, Cleveland Clinic, Cleveland, Ohio 44195.
Purpose:
Flood sources of (57)Co are commonly used for quality control of gamma cameras. Flood uniformity may be affected by the contaminants (56)Co and (58)Co, which emit higher energy photons. Although vendors specify a maximum combined (56)Co and (58)Co activity, a convenient test for flood source purity that is feasible in a clinical environment would be desirable.
Methods:
Both (56)Co and (58)Co emit positrons with branching 19.6% and 14.9%, respectively. As is known from (90)Y imaging, a positron emission tomography (PET) scanner is capable of quantitatively imaging very weak positron emission in a high single-photon background. To evaluate this approach, two (57)Co flood sources were scanned with a clinical PET/CT multiple times over a period of months. The (56)Co and (58)Co activity was clearly visible in the reconstructed PET images. Total impurity activity was quantified from the PET images after background subtraction of prompt gamma coincidences.
Results:
Time-of-flight PET reconstruction was highly beneficial for accurate image quantification. Repeated measurements of the positron-emitting impurities showed excellent agreement with an exponential decay model. For both flood sources studied, the fit parameters indicated a zero intercept and a decay half-life consistent with a mixture of (56)Co and (58)Co. The total impurity activity at the reference date was estimated to be 0.06% and 0.07% for the two sources, which was consistent with the vendor's specification of <0.12%.
Conclusions:
The robustness of the repeated measurements and a thorough analysis of the detector corrections and physics suggest that the accuracy is acceptable and that the technique is feasible. Further work is needed to validate the accuracy of this technique with a calibrated high resolution gamma spectrometer as a gold standard, which was not available for this study, and for other PET detector models.
More Related Videos
10:31Detection and Recovery of Palladium, Gold and Cobalt Metals from the Urban Mine Using Novel Sensors/Adsorbents Designated with Nanoscale Wagon-wheel-shaped Pores
Published on: December 6, 2015
08:40Synthesis of Metal Nanoparticles Supported on Carbon Nanotube with Doped Co and N Atoms and its Catalytic Applications in Hydrogen Production
Published on: December 6, 2021
Related Concept Videos
Precipitation and Co-precipitation
Atomic Emission Spectroscopy: Lab